Cyclone | Long Pathlength Gas Cell
The cyclone™ is a standard long pathlength gas cell accessory designed with Borosilicate glass body and gold mirrors which are protected. They have been optimized using the latest optical modelling tools to give maximum signal-to-noise.
Configurable long pathlength
There are made to a specified pathlength within the ranges described below. Each of the four bodies permits a different range and has a different internal volume.
Adjustable mirror
There is adjustable mirror carriage which can be fitted to Cyclone™ cells to enable the user to vary the pathlength throughout the specified range.
Features
Wide pathlength range (1m - 10m)
Vacuum to 15 p.s.i. operation
Ambient temperature operation
Borosilicate glass body
Anodised components
Gold mirrors (protected)
Viton or Kalrez 'O' ring seals
KBr, CaF2 or ZnSe windows
Purgeable transfer optics box
Benchmark series baseplate mounting
Optional Features
Additional mirror carriage assemblies
Vacuum / gas inlet & outlet taps
Pressure gauge
Desiccant storage caps
Purge bellows
Flexible specification long pathlength gas cells | Cyclone™
These cells are available with glass or metal cell bodies and can be set to a range of pathlengths.
Model
C2
C5 or T5
C10 or T10
Pathlength settings (m)
0.5-2.5
in 0.5 m steps
1.0-8.0
in 1.0 m steps
2.1-10.6
in 1.06 m steps
Base pathlength (mm)
125
250
264
Number of passes
4-20
4-32
8-40
Volume (L)
0.19
1.33
2.60
Which pathlength should I choose?
The absorbance of a gas depends on the distance travelled by the IR light beam through the gas sample. The relationship between absorbance, A, concentration, C, and pathlength, L, is given by Beer’s Law:
A = -log10 (I/I0) = a.C.L
Atmospheric concentrations of gases are usually expressed in C.L units of ppm.m – the number of molecules that would be encountered by the infrared beam across a 1.0 m path.
A gas at 0.1 ppm atmospheric concentration will absorb as much IR light over a pathlength of 100 m as would the same gas at 1 ppm over 10 meters or at 10 ppm over 1 meter.
To achieve optimal results, the pathlength of cell should be chosen to give absorbance values within the spectrometer’s linear range for a given concentration. The following table may be taken as a guide: